标准编号:ISO/ASTM 51650:2013

中文名称:醋酸纤维素剂量测定系统的使用规程

英文名称:Practice for use of a cellulose triacetate dosimetry system

发布日期:2013-06

标准范围

1.1 This is a practice for using a cellulose triacetate (CTA) dosimetry system to measure absorbed dose in materials irradiated by photons or electrons in terms of absorbed dose to water. The CTA dosimetry system is classified as a routine dosimetry system.1.2 The CTA dosimeter is classified as a type II dosimeter on the basis of the complex effect of influence quantities on its response (see ASTM Practice E2628).1.3 This document is one of a set of standards that provides recommendations for properly implementing dosimetry in radiation processing, and describes a means of achieving compliance with the requirements of ASTM E2628 “Practice for Dosimetry in Radiation Processing” for a CTA dosimetry system. It is intended to be read in conjunction with ASTM E2628.1.4 This practice covers the use of CTA dosimetry systems under the following conditions:1.4.1 The absorbed dose range is 10 kGy to 300 kGy.1.4.2 The absorbed-dose rate range is 3 Gy/s to 4×10(10) Gy/s (1).(2)1.4.3 The photon energy range is 0.1 to 50 MeV.1.4.4 The electron energy range is 0.2 to 50 MeV.1.5 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.